Alarm:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Analogové výstupy:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Automatické čištění:
Ano
Certifikace:
CE compliant / ETL certified
Demineralised water:
Consumption: 140 mL/analysis (hour)
Digitální výstupy:
Optional: Modbus (TCP/IP, RS485)
Doba cyklu :
Standard measurement cycle time for both TN & TP: 60 minutes
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Hmotnost:
25 kg
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interference :
TN: The main interferences are Br- and I-. When the amount of I- is 2.2 fold of the amount of TN, or the amount of Br- is 3.4 fold of the amount of TN, this will interfere on the test results. Dissolved organic matter, surfactants and Chromium (VI) interfere. Various inorganic substances not normally found in natural water, such as Chlorite [ClO₂-] and Chlorate [ClO₃-], may interfere.
TP: Arsenic (V), Chromium (VI), Copper (II) > 10 mg/L, Iron (III) > 10 mg/L, Sulphide > 2 mg/L, and Vanadium (V), Silica > 60 mg/L. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Jaký je obsah balení:
EZ7653 Total Nitrogen & Total Phosphate Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 3 x empty 2.5L Reagent Containers with Fittings (HCl, NaOH and H₂SO₄ Solution) and 3 x empty 2.5L Glass Reagent Containers with Fittings (Colour, Reductor & Persulphate Solution)
Kalibrace:
Automatic, 2-point; frequency freely programmable
Materiál:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metoda měření:
TN: UV photometric measurement at 220 nm after persulphate digestion in alkaline medium, based on APHA 4500-NO3 (B)
TP: Colorimetric measurement at 700 nm using ascorbic acid reduction and molybdate colour solution after persulphate digestion in acidic medium, based on APHA 4500-P
Mez detekce:
TN: ≤ 0.5 mg/L
TP: ≤ 0.025 mg/L
Napájení:
230 VAC, 50/60 Hz
120 VAC, 50/60 Hz
Max. power consumption: 440 VA
Odtok:
Atmospheric pressure, vented, min. Ø 32 mm
Okolní teplota:
10 - 30 °C, ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Parametr:
Total Nitrogen (TN), Total Phosphorus (TP)
Počet proudů vzorku:
1, 2, 4, or 8
Požadavky na vzorek:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Přesnost:
Better than 3% (TN) and 2% (TP) full scale range for standard test solutions
Průtoková rychlost:
100 - 300 mL/min
Reagent Requirements:
Keep between 10 - 30 °C
Rozměry (V x Š x H):
690 mm x 465 mm x 330 mm
Rozsah měření:
0.5 - 20 mg/L TN; 0.025 - 5 mg/L TP
Teplota vzorku:
10 - 30 °C
Tlak vzorku:
By external overflow vessel
Třída ochrany:
Analyser cabinet: IP55 / Panel PC: IP65
Validation:
Automatic; frequency freely programmable
Záruka: