Alarm:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Analogové výstupy:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Certifikace:
CE compliant / UL certified
Demineralised water:
For rinsing / dilution
Digitální výstupy:
Optional: Modbus (TCP/IP, RS485)
Doba cyklu :
20 - 30 minutes
40 minutes (+ Total Hardness Ca)
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Hmotnost:
25 kg
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interference :
Hardness: Some metal ions interfere by causing fading or indistinct end points or by stoichiometric consumption of EDTA. Suspended or colloidal organic matter also may interfere with the end point. Large amounts of colour and turbidity interferes. Fats, oil, proteins, surfactants and tar.
Alkalinity: Soaps, oily matter, suspended solids or precipitates may coat the glass electrode and cause a sluggish response. Allow additional time between titrant additions to let the electrode come to equilibrium or clean the electrodes occasionally.
Jaký je obsah balení:
EZ5001 Total / Free Alkalinity Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Bracket, 1 x pH Electrode, Routine S7 Connection and 1 x empty 10L Reagent Container with Fittings (HCl Solution)
Kalibrace:
N.A.
Materiál:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metoda:
Titrimetric
Metoda měření:
Acid-base titration (alkalinity only); Calmagite EDTA titration with LED dipping probe (combined)
Mez detekce:
≤ 10 mg/L
Napájení:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Odtok:
Atmospheric pressure, vented, min. Ø 32 mm
Okolní teplota:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Parametr:
Alkalinity, total; Alkalinity, free
Počet proudů vzorku:
1, 2, 4 or 8
Požadavky na vzorek:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Přesnost:
Better than 2% full scale range for standard test solutions
Průtoková rychlost:
100 - 300 mL/min
Reagent Requirements:
Keep between 10 - 30 °C
Rozměry (V x Š x H):
690 mm x 465 mm x 330 mm
Rozsah měření:
100 - 5000 mg/L CaCO3
Optional:
10 - 500 mg/L CaCO3
25 - 1250 mg/L CaCO3
50 - 2500 mg/L CaCO3
Internal dispenser dilution (max. factor 100)
100 - 5000 mg/L CaCO₃
Teplota vzorku:
10 - 30 °C
Tlak vzorku:
By external overflow vessel
Třída ochrany:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Validation:
Automatic; frequency freely programmable
Záruka:
24 měsíců