AMTAX indoor sc MB2(0.05-20 mg/l) filter probe (230VAC/50HZ)

Č. produktu: LXV421.99.11011
CZK Cena: Kontaktujte nás

High precision process-instrument for the determination of NH4 content in water, waste water and activated sludge. The measurement is operating a gas sensitive electrode with unscrewable membrane cap.
The sample preparation is done by an integrated sample preparation probe with 5 m heated hose. The filtration probe is operating at 230V. Evaluation and operation with SC 1000 Controller. The analyser is equipped with an insulated enclosure for indoor installation. The sample can be analysed before any degradation. The ammonium in the sample is first converted to gaseous ammonia. Only the NH3 gas passes through the gas-permeable membrane of the electrode and is detected. This method guarantees a wide measuring range and is less susceptible to cross-sensitivity methods that make use of an ion-selective electrode (ISE). in situ membrane filtration; filter modules are exchangeable.

  • Automatické čištění a kalibrace
  • Rozsáhlá samodiagnostika
  • Možnost instalace na stěnu, zábradlí a stojan
  • Jednoduchá obsluha a provoz
  • S integrovanou filtrační sondou